Salta al contenuto principale
Unisciti alla Community delle riparazioni - Crea un account

The ResMed VPAP Auto 25, also known as the ResMed S8 VPAP Auto 25, is a bilevel positive airway pressure (BiPAP) device used to treat sleep apnea. This page provides repair and maintenance information, including common issues and component-level guides.

Receiving system error 1018

Receiving System error 1018 to call service. Unplugged and waited 5-10 minutes, machine still doesn’t work.

Thank you,

Camille Kutis

Rispondi a questa domanda Anch'io ho questo problema

Questa è una buona domanda?

Punteggio 0
Aggiungi un commento

1 Risposta

Risposta Più Utile

@corkie The following message is displayed on the LCD: "SYSTEM ERR 1018 Call service!"

The error is a Flow sensor fault. The solution is to Replace flow sensor. You can download the service manual from this page ResMed VPAP Auto 25 to work on your CPAP. Ultimately you do want somebody (preferably a BioMed person) look it over after you do your repair.

Questa risposta è stata utile?

Punteggio 1
Aggiungi un commento

Aggiungi la tua risposta

Camille Kutis sarà eternamente grato.
Visualizza Statistiche:

Ultime 24 Ore: 0

Ultimi 7 Giorni: 6

Ultimi 30 Giorni: 18

Tutti i Tempi: 736