Salta al contenuto principale
Unisciti alla Community delle riparazioni - Crea un account

Autosampler Programming (For Old & New Versions)

    1. Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: passo 1, immagine 1 di 2 Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: passo 1, immagine 2 di 2
      • 1. Switch battery "ON".

      • 2. Autosampler: Push "ON" button.

      • It will say ...STANDBY...

      Chiedi a FixBot

      Aggiungi Commento

    2. Autosampler Programming  (For Old & New Versions), Select Program Mode: passo 2, immagine 1 di 2 Autosampler Programming  (For Old & New Versions), Select Program Mode: passo 2, immagine 2 di 2
      • 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM” button to select "PROGRAM".

      • 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.

      Chiedi a FixBot

      Aggiungi Commento

    3. Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: passo 3, immagine 1 di 3 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: passo 3, immagine 2 di 3 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: passo 3, immagine 3 di 3
      • 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button.

      • 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button.

      • 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.

      Chiedi a FixBot

      Aggiungi Commento

    4. Autosampler Programming  (For Old & New Versions), Starting Sampling: passo 4, immagine 1 di 3 Autosampler Programming  (For Old & New Versions), Starting Sampling: passo 4, immagine 2 di 3 Autosampler Programming  (For Old & New Versions), Starting Sampling: passo 4, immagine 3 di 3
      • 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED".

      • 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1".

      • 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).

      • It will take some time before it starts sampling.

      • Eventually screen will show: " BOTTLE 1 AFTER 20 PULSES".

      Chiedi a FixBot

      Aggiungi Commento

Linea Traguardo

Annulla: non ho completato questa guida.

Un'altra persona ha completato questa guida.

Brooke Mason

Membro da: 05/29/19

2.958 Reputazione

17 guide realizzate

Team

Open-Storm Membro di Open-Storm

Community

14 Membri

266 guide realizzate

0Commenti sulla guida

Aggiungi Commento

Visualizza statistiche:

Ultime 24 ore: 0

Ultimi 7 giorni: 4

Ultimi 30 giorni: 29

Tutti i Tempi: 510