LITHOSTAR Multiline

' Siemens AG 1996 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.

SP

System

Errors by error codes

Troubleshooting Instructions

Print No.: RXL2-120.072.03.04.02

English

Doc. Gen. Date:

06.02

ChapterPageRevision
allall04

Table Caption: This table indicates that revision 04 applies to all chapters and pages.

Document revision level

The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.

Please contact your local Siemens office to order current revision levels.

Disclaimer

The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.

Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.

Rev. 04

06.02

0

0

Table Caption: This table is a table of contents for a technical manual, likely related to medical imaging equipment, detailing the different controllers (Motor, Communication, CPU) and their associated error codes and specific system components.

Table Caption: This table is an index or table of contents that lists topics related to boot software, error messages, and operating status displays, along with their corresponding page numbers. It covers various components and error conditions of a PC system.

General errors

Error number

  • ↓ Meaning:
  • 01 PROM_error
  • 02 INT_RAM_error
  • 03 EXT_RAM_error
  • 04 TIMER_0_error
  • 05 TIMER_1_error
  • 06 KERN_TEST_error
  • 07 SFR_error
  • 09 WATCHDOG_error

ZR am1 (Lifting column right)

Error number

  • ↓ Meaning:
  • 10 LM_628_error
  • 11 Control_error
  • 12 Position accuracy not reached in stated stabilizing time
  • 13 Motor not initialized, drive command received
  • 14 Motor moves without actuation
  • 15 Motor does not move when actuated
  • 16 Speed not reached or exceeded
  • 17 Motor moves in wrong direction
  • 18 Motor unsuccessfully actuated three times
  • 19 Faulty initialization of the ASIC

ZL am2 (Lifting column left)

Error number

  • ↓ Meaning:
  • 20 LM_628_error
  • 21 Control_error
  • 22 Position accuracy not reached in stated stabilizing time
  • 23 Motor not initialized, drive command received
  • 24 Motor moves without actuation
  • 25 Motor does not move when actuated
  • 26 Speed not reached or exceeded
  • 27 Motor moves in wrong direction
  • 28 Motor actuated unsuccessfully three times
  • 29 Faulty initialization of the ASIC

1