Domande popolari nel forum
Fai una domandaNon ci sono domande. Puoi essere il primo a porre una domanda!
Documenti
Background
The Bard Site Rite 6 is a portable, hospital-grade ultrasound system designed to provide real-time guidance for the insertion of needles and catheters in vascular structures. The system can also be used for the ultrasound imaging of internal organs. While no longer available for sale on the official BD website, the Bard Site Rite 6 can still be purchased online through third-party medical retailers.
The Bard Site Rite 6 system comes pre-installed with version 1.1.3 of BD's Sherlock 3CG Software. The Sherlock 3CG software is programmed with a real-time needle guiding interface which provides users with a clear visualization of vessels. The software is designed to help clinicians make successful needle insertions, reducing the risk of arterial punctures, hematomas, or pneumothorax.
In 2018, a malfunction was discovered in the Bard Site Rite 6 system's pre-installed Sherlock 3CG software when it failed to properly track the PICC (Peripherally Inserted Central Catheter) magnet during catheter insertion, leading to an improperly positioned PICC.
Identification
The Bard Site Rite 6 ultrasound system has a square, box-like appearance. The device is encased in a plastic shell that surrounds the large screen on the front of the system. A handle is built into the top of the shell, and a base is built into the bottom.
Below the screen are six buttons, with the rightmost one being the power button.
In the center of the right side of the system (screen facing you) is a large, rectangular outlet which is used to plug in the probe holster.
Identifiable Traits:
- The BARD logo is found on the front, top left-hand side of the device.
- A Site~Rite 6 logo sticker is located on the front, top right-hand side of the device.
- Another Site~Rite 6 logo sticker is located on the back right-hand side of the system.
- On the side of the probe is a printed label warning that it may only be used with the Bard Site Rite 6 system.
- A variation of the Bard Site Rite 6's reference number (97770000) can be found on a sticker on the back of the device, as well as printed onto the side of the probe.
Technical Specifications
Scanner
- Unit Dimensions:
- Width: 12 in (30.5 cm)
- Height: 13 in (33.02 cm)
- Depth: 5 in (12.7 cm)
- Monitor Size: 12.1 in (30.7 cm) diagonal
- Weight: 10 lbs. (4.5 kg)
- Power Source: A/C adapter, Internal and External DC Battery Pack
- Power Consumption: 84 Watts Maximum
Power Supply
- A/C Adapter Specifications:
- Input Voltage: 100-200 VAC, 50/60 Hz.
- Input Current: 2 Amps Maximum
- Output Voltage: 15 VDC
- Output Current: 6 Amps Maximum
- Internal Battery Pack Specifications:
- Battery Chemistry: Lithium Ion
- Nominal Output Voltage: 10.8 VDC
- Output Current: 6 Amps Maximum
- Output Power at Full Charge: 52 Wh
- System Run Time on Full Charge: 1.0 Hours
- Battery Charge Time: 1.75 Hours
- Combination A/C Adapter Auxiliary Battery Specifications:
- Battery Chemistry: Lithium Ion
- Input Voltage: 100-240 VAC, 50/60 Hz
- Input Current: 1.7 Amps Maximum
- A/C Adapter Output Voltage: 15 VDC
- A/C Adapter Output Current: 7 Amps Maximum
- Nominal Battery Output Voltage: 10.8 VDC
- Battery Output Current: 6 Amps Maximum
- Output Power at Full Charge: 95 Wh
- System Run Time on Full Charge: 2.5 Hours
- Battery Charge Time: 3 Hours
Probe
- Frequency: 5-10 MHz
- Elevation Focus: 1.8 cm
- Maximum Scan Depth: 2.36 in 6.0 cm
- Scan Width: 0.75 in (1.9 cm)
- Lateral Foci:
- Image Depth: 1.5 cm; Focal Depth: 0.6 cm
- Image Depth: 3.0 cm; Focal Depth: 1.5 cm
- Image Depth: 4.5 cm; Focal Depth: 3.0 cm
- Image Depth: 6.0 cm; Focal Depth: 5.0 cm
Operating Temperature and Storage Conditions
- Operating Temperature: 59°F to 100°F (15°C to 38°C)
- Storage Temperature: 50°F to 100°F (10°C to 38°C)
- Operating Humidity: 5% to 85% Relative Humidity (non-condensing)
- Storage Humidity (packaged): 5% to 95% Relative Humidity (non-condensing)
- Storage Humidity (unpackaged): 5% to 85% Relative Humidity (non-condensing)