The Water in the Humidifier Will Not Heat Up
The water in the ventilator’s humidifier remains cold even after the device has been turned on for a while.
Humidifier is defective
Test the Ventilator with another humidifier to confirm whether or not the
current one is defective. If it is defective, return it to the manufacturer
Power Board is Defective
Test to see if the power board is defective by considering these other
issues: there is no running noise and displays will not show up or when in
automatic mode, the device cannot be switched on by a breath being
taken in. If these 2 other issues are also occurring, replace the power
board (see Weinmann VENTImotion and VENTIlogic Power Board Replacement) and return it to the
manufacturer.
There are No Displays, or they are Faulty
The displays shown on the screen of the ventilator are inaccurate or nothing is showing.
Display is defective
Replace Display (see Weinmann VENTImotion and VENTIlogic Display Board Replacement)
Cable came loose or is defective
Check all cable connections in device; replace if necessary
Control Board is defective
Test to see if the control board is defective by considering these other
issues: Tolerance of therapy pressures is >0.6 hPa after 1 minute, in
automatic mode or device cannot be switched on by a breath being taken
in. If these 2 other issues are also occurring, replace the control board
(see Weinmann VENTImotion and VENTIlogic Control Board Replacement)
There is No Running Noise and the Displays Will Not Show Up
When the ventilator is turned on, the screen is black and there is no noise coming from the machine.
There is no power supply
Check connection of Power Supply cable in the device socket. Check the
source of voltage with another device (for example; a lamp). Replace
power supply cable if necessary.
The main supply unit is defective
Replace main supply unit (see VENTImotion & VENTIlogic Service and Repair Instructions , page 33)
The cable came loose or is defective
Check all cable connections in the device; replace if necessary.
The control board is defective
Test to see if the control board is defective by considering these other
issues: Tolerance of therapy pressures is >0.6 hPa after 1 minute, in
automatic mode or device cannot be switched on by a breath being taken
in. If these 2 other issues are also occurring, replace the control board
(see Weinmann VENTImotion and VENTIlogic Control Board Replacement).
The power board is defective
Test to see if the power board is defective by considering these other
issues: the humidifier will not heat up or the device cannot be switched on
by a breath being taken in. If these 2 other issues are also occurring,
replace the power board (see Weinmann VENTImotion and VENTIlogic Power Board Replacement) and
return it to the manufacturer.
Device Cannot Be Turned on using the Film Keypad
Device will not power on when the power button on the film keypad is pressed.
The ribbon cable of the film keypad has slipped out of the connection
Check whether the ribbon cable is correctly connected to the display
board.
The Fascia Film is defective
Replace fascia film (see Weinmann VENTImotion and VENTIlogic Fascia Film Replacement)
The Power Supply Cable is Not Providing Power
The power supply cable is defective
Replace the Power Supply Cable (see Weinmann VENTImotion and VENTIlogic Power Board Replacement)