Help! Ohmeda 3740 Biox repair required.

Hi

I have this unit in great condition except it's stopped working ! The green power light is illuminated when plugged into mains charging but nothing else works.

An ideas or even better, an offer of repair ?

I live in Kent UK

Best wishes.

John

Rispondi a questa domanda Anch'io ho questo problema

Questa è una buona domanda?

Punteggio 0
Aggiungi un commento

1 Risposta

Risposta Più Utile

@resusman take alook at the maintenencae manual for your pulse oximeter Ohmeda Biox 3740 - Operation and Maintenance Manual.pdf page 49 suggest that this is either a probe failure or a circuit failure. More links for a service guide for this device can be found at the bottom of this page Datex-Ohmeda 3740