Salta al contenuto principale

Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement

Cosa ti serve

  1. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: passo 1, immagine 1 di 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: passo 1, immagine 2 di 2
    • Unplug the AC power connection cable.

  2. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 2, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 2, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 2, immagine 3 di 3
    • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

    • Always grasp the hose by the connector. Do not pull on the hose itself.

    • Pull the connector away from the connector port.

  3. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 3, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 3, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 3, immagine 3 di 3
    • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

    • Always grasp the cable by the connector. Do not pull on the cable itself.

    • Pull the connector away from the connector port.

  4. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 4, immagine 1 di 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 4, immagine 2 di 2
    • Grasp the temperature probe and pull it up to remove it from the monitor.

  5. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 5, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 5, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 5, immagine 3 di 3
    • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

  6. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 6, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 6, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 6, immagine 3 di 3
    • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

  7. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 7, immagine 1 di 1
    • Remove the flathead screw on the USB networking door.

    • Loosen the captive Phillips #2 screw securing the monitor to the stand.

  8. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 8, immagine 1 di 1
    • Holding the monitor securely, open the USB networking door.

    • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

  9. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 9, immagine 1 di 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 9, immagine 2 di 2
    • Detach any accessory USB cables from USB ports on the monitor.

  10. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 10, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 10, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 10, immagine 3 di 3
    • Slide the monitor off of the stand tray.

Conclusione

To reassemble your device, follow these instructions in reverse order.