Salta al contenuto principale

Questa guida ha delle modifiche più recenti. Passa all'ultima versione non verificata.

Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement

Cosa ti serve

  1. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: passo 1, immagine 1 di 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: passo 1, immagine 2 di 2
    • Unplug the AC power connection cable.

  2. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 2, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 2, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 2, immagine 3 di 3
    • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

    • Always grasp the hose by the connector. Do not pull on the hose itself.

    • Pull the connector away from the connector port.

  3. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 3, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 3, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 3, immagine 3 di 3
    • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

    • Always grasp the cable by the connector. Do not pull on the cable itself.

    • Pull the connector away from the connector port.

  4. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 4, immagine 1 di 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 4, immagine 2 di 2
    • Grasp the temperature probe and pull it up to remove it from the monitor.

  5. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 5, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 5, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 5, immagine 3 di 3
    • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

  6. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 6, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 6, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 6, immagine 3 di 3
    • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

  7. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 7, immagine 1 di 1
    • Remove the flathead screw on the USB networking door.

    • Loosen the captive Phillips #2 screw securing the monitor to the stand.

  8. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 8, immagine 1 di 1
    • Holding the monitor securely, open the USB networking door.

    • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

  9. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 9, immagine 1 di 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 9, immagine 2 di 2
    • Detach any accessory USB cables from USB ports on the monitor.

  10. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 10, immagine 1 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 10, immagine 2 di 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: passo 10, immagine 3 di 3
    • Slide the monitor off of the stand tray.

Conclusione

To reassemble your device, follow these instructions in reverse order.

Sam Goldheart

Membro da: 10/18/12

465.611 Reputazione

538 Guide realizzate

Team

iFixit Membro di iFixit

Staff

135 Membri

16.731 Guide realizzate

0 Commenti

Aggiungi Commento

Visualizza Statistiche:

Ultime 24 Ore: 0

Ultimi 7 Giorni: 1

Ultimi 30 Giorni: 3

Tutti i Tempi: 861