Unplug the AC power connection cable.
  • Unplug the AC power connection cable.

Aggiungi Commento

Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.
  • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

    • Always grasp the hose by the connector. Do not pull on the hose itself.

  • Pull the connector away from the connector port.

Aggiungi Commento

  • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

    • Always grasp the cable by the connector. Do not pull on the cable itself.

  • Pull the connector away from the connector port.

Aggiungi Commento

  • Grasp the temperature probe and pull it up to remove it from the monitor.

Aggiungi Commento

  • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

Aggiungi Commento

  • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

Aggiungi Commento

  • Remove the flathead screw on the USB networking door.

  • Loosen the captive Phillips #2 screw securing the monitor to the stand.

Aggiungi Commento

  • Holding the monitor securely, open the USB networking door.

    • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

Aggiungi Commento

  • Detach any accessory USB cables from USB ports on the monitor.

Aggiungi Commento

  • Slide the monitor off of the stand tray.

Aggiungi Commento

  • Place the monitor upside down on a clean surface.

  • Insert a coin into the slot and push to open.

    • Choose a coin that fits comfortably into the slot.

  • Remove the battery cover.

Aggiungi Commento

  • Use the plastic label to remove the battery from its recess.

Aggiungi Commento

  • Open the communications door to access the lower left-hand rear housing screw.

  • Remove the four Phillips #2 screws (labeled 14 in the service manual) from the rear housing.

Aggiungi Commento

  • With the handle/alert bar facing you, begin to lift the rear housing from the left side, holding the right side securely.

  • Do not attempt to separate the rear housing from the monitor, it is still attached with several cables.

Aggiungi Commento

  • Support the rear housing while you disconnect the cables in the following steps.

Aggiungi Commento

  • Disconnect the fan cable (labeled connector A in the service manual) from its socket in the power supply.

Aggiungi Commento

  • Unplug the small USB connector from its socket (labeled J4 in the service manual).

Aggiungi Commento

  • Disconnect the communications power cable from its socket (labeled J50 in the service manual).

  • Only pull on the connector itself and not the socket. This connection is very tight, work the connector slowly and gently.

Aggiungi Commento

  • Unplug the large USB connector from its socket (labeled J8 in the service manual).

Aggiungi Commento

  • Press the tab on the ethernet cable and unplug it from its socket (labeled J9 in the service manual).

Aggiungi Commento

  • The following connector requires two hands to disconnect. Flip the rear case over, and lay it on its back to gain access to the power supply cable.

Aggiungi Commento

  • The power supply cable is secured by an interlocking connector that must be held open to unplug the connector.

  • Insert the flat end of a spudger down into the power supply cover between the socket clip and the power supply connector.

Aggiungi Commento

  • Grasp the power supply cable firmly and lift both it and the spudger from the channel in the power supply housing to unplug the connector.

Aggiungi Commento

  • Remove the four T10 Torx screws from the power supply cover.

Aggiungi Commento

  • Remove the power supply cover.

Aggiungi Commento

  • Unplug the AC power harness connectors from the IEC connector with a pair of pliers.

  • During reassembly connect green to ground, blue to line (L), and brown to neutral (N).

Aggiungi Commento

  • Disconnect the ground connector from the ground plate.

  • During reassembly plug the ground cable from the IEC connector onto the ground plate in the position closest to the IEC connector. Ensure that the green ground wires are routed over the blue and brown wires.

Aggiungi Commento

  • Use the flat end of a spudger to lift the power supply up off the posts in the rear housing.

  • Lift and remove the power supply from the rear housing.

    • Be very careful not to touch the capacitor leads or any exposed solder joints on the back of the power supply. Only handle the board by the edges.

  • During reassembly ensure that the rubber grommet on the AC power harness is in the slot in the power-supply cover.

  • Ensure that the green ground wires are secured under the power-supply cover.

Aggiungi Commento

Conclusione

To reassemble your device, follow these instructions in reverse order.

Un'altra persona ha completato questa guida.

Sam Lionheart

Membro da: 18/10/2012

333.484 Reputazione

534 Guide realizzate

Team

iFixit Membro di iFixit

Community

107 Membri

6.510 Guide realizzate

Aggiungi Commento

Visualizza Statistiche:

Ultime 24 Ore: 0

Ultimi 7 Giorni: 1

Ultimi 30 Giorni: 6

Tutti i Tempi: 476