Philips Respironics V60 Top Cover Replacement

Attrezzature medicali
Né iFixit né questa guida di riparazione sono approvati dal produttore del dispositivo medicale o affiliati ad esso.

Cosa ti serve

  1. Philips Respironics V60 Top Cover Replacement, shut down the ventilator: passo 1, immagine 1 di 2 Philips Respironics V60 Top Cover Replacement, shut down the ventilator: passo 1, immagine 2 di 2
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. Philips Respironics V60 Top Cover Replacement: passo 4, immagine 1 di 1
    • Disconnect the internal battery from the power harness connector.

  3. Philips Respironics V60 Top Cover Replacement, Top Cover: passo 5, immagine 1 di 1
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

  4. Philips Respironics V60 Top Cover Replacement: passo 6, immagine 1 di 1
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

Conclusione

To reassemble your device, follow these instructions in reverse order.

Rie Kuraoka

Membro da: 01/23/20

886 Reputazione

7 Guide realizzate

Team

Cal Poly, Team S7-G10, Paton Spring 2020 Membro di Cal Poly, Team S7-G10, Paton Spring 2020

CPSU-PATON-S20S7G10

2 Membri

9 Guide realizzate

0 Commenti

Aggiungi Commento

Visualizza Statistiche:

Ultime 24 Ore: 0

Ultimi 7 Giorni: 0

Ultimi 30 Giorni: 4

Tutti i Tempi: 266