  1. Philips BiPAP S/T Humidifier Replacement, Humidifier: passo 1, immagine 1 di 1
    • Remove the Water Chamber Assembly.

    • Gently squeeze the latch on the Flip Lid Assembly to release it.

    • Lift the Flip Lid Assembly.

    • Pull the Humidifier Tank Assembly out of the Humidifier.

  2. Philips BiPAP S/T Humidifier Replacement: passo 2, immagine 1 di 1
    • Remove the Tank Top Seal from the Flip Lid Assembly.

    • If this is the source of the error, put in a new one.

  3. Philips BiPAP S/T Humidifier Replacement: passo 3, immagine 1 di 2 Philips BiPAP S/T Humidifier Replacement: passo 3, immagine 2 di 2
    • Remove the Dry Box Seal.

    • If this is the source of error, place this part with a new one.

    • When replacing, ensure that the narrow end of the seal is inserted into the Dry Box.

  4. Philips BiPAP S/T Humidifier Replacement: passo 4, immagine 1 di 2 Philips BiPAP S/T Humidifier Replacement: passo 4, immagine 2 di 2
    • Remove the Dry Box Assembly.

    • Insert a flat blade screwdriver into the hole located on the back of the Humidifier and lightly press inward to release the Dry Box Assembly.

    • Do not press firmly on the screwdriver as damage may occur.

    • Remove the Inlet Seal from the Dry Box Assembly

  5. Philips BiPAP S/T Humidifier Replacement: passo 5, immagine 1 di 3 Philips BiPAP S/T Humidifier Replacement: passo 5, immagine 2 di 3 Philips BiPAP S/T Humidifier Replacement: passo 5, immagine 3 di 3
    • Remove the Flip Lid Assembly.

    • Use a screwdriver to push in on the latches that secure the Flip Lid Assembly to the Humidifier Top Housing.

    • Bend the Flip Lid Assembly completely backwards until it is removed.

    • For Heated Tube devices, remove the wire harness from the base.

  6. Philips BiPAP S/T Humidifier Replacement: passo 6, immagine 1 di 1
    • Use a T8 Torx screwdriver to remove the four #4 x 1/2" screws that secure the Top Housing to the Humidifier Bottom Housing

  7. Philips BiPAP S/T Humidifier Replacement: passo 7, immagine 1 di 1
    • Slide the Outside Cover out of the Bottom Housing.

  8. Philips BiPAP S/T Humidifier Replacement: passo 8, immagine 1 di 1
    • Remove the Left Side Panel.

    • Use a T15 Torx screwdriver to remove the four #6 x 1/4" screws that secure the Bottom Housing to the Lower Base Assembly.

  9. Philips BiPAP S/T Humidifier Replacement: passo 9, immagine 1 di 2 Philips BiPAP S/T Humidifier Replacement: passo 9, immagine 2 di 2
    • Remove the Heater Plate Assembly.

    • The Heater Plate Assembly is secured by two small tabs.

    • Once replaced, secure the Bottom Housing to the Lower Base using the four #6 x 1/4" screws.

    • Be sure the Heater Plate Spring is properly seated under the Heater Plate.

Conclusione

To reassemble your device, follow these instructions in reverse order.

Courtney Huitt

Membro da: 04/07/20

169 Reputazione

1 Guida realizzata

Team

Cal Poly, Team S7-G3, Paton Spring 2020 Membro di Cal Poly, Team S7-G3, Paton Spring 2020

CPSU-PATON-S20S7G3

3 Membri

3 Guide realizzate

0 Commenti

Aggiungi Commento

Visualizza Statistiche:

Ultime 24 Ore: 0

Ultimi 7 Giorni: 2

Ultimi 30 Giorni: 5

Tutti i Tempi: 281