Opening the Column in the Hamilton Galileo

Cosa ti serve

  1. Opening the Column in the Hamilton Galileo, Opening the Column in the Hamilton Galileo: passo 1, immagine 1 di 1
    • Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.

    • Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).

  2. Opening the Column in the Hamilton Galileo: passo 3, immagine 1 di 1
    • Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.

  3. Opening the Column in the Hamilton Galileo: passo 4, immagine 1 di 2 Opening the Column in the Hamilton Galileo: passo 4, immagine 2 di 2
    • Without removing any electrical connections, open the left side of the column as wide as practical.

    • This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.

    • Make sure no electrical connections are touching the column.

    • Make sure all green/brown grounding (earthing) cables are securely in place.

Conclusione

To reassemble your device, follow these instructions in reverse order.