    • Set the gas supply to 5 L/min.

  1. How to Test the Valve System on a Fisher Paykel Neopuff: passo 2, immagine 1 di 1
    • Completely close the Peak Inspiratory Pressure control and the Max Pressure control by turning both knobs completely clockwise.

  2. How to Test the Valve System on a Fisher Paykel Neopuff: passo 3, immagine 1 di 1
    • Close the bleed valve and check that the gauge reads at least 60 cmh2O.

    • Set the gas supply to 15 L/min.

  3. How to Test the Valve System on a Fisher Paykel Neopuff: passo 5, immagine 1 di 1
    • Close the bleed valve and check that the gauge reads no higher than 80 cmH2O.

    • Set the gas supply to 10 L/min.

    • Close the bleed valve and turn the Max Pressure Relief valve until the manometer reads 40 cmh2O.

    • Check that the manometer needle rises and falls smoothly.

    • Gently rotate the Max Pressure Relief cover over the Max Pressure Relief knob.

  4. How to Test the Valve System on a Fisher Paykel Neopuff: passo 10, immagine 1 di 1
    • Reset the peak inspiratory pressure to 20 cmh2O and turn off the gas flow. Testing is now complete. If the F&P Neopuff/Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers).

Conclusione

To reassemble your device, follow these instructions in reverse order.

Un'altra persona ha completato questa guida.

Team

Cal Poly, Team S1-G6, Zimmerman Spring 2020 Membro di Cal Poly, Team S1-G6, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G6

4 Membri

27 Guide realizzate

0 Commenti

Aggiungi Commento

Visualizza Statistiche:

Ultime 24 Ore: 1

Ultimi 7 Giorni: 2

Ultimi 30 Giorni: 5

Tutti i Tempi: 497