Please refer to page 7 of the device’s [document|6918|Technical Manual] for source information.

    • Adjust the gas flow to 10 L/min.

  1. Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.
    • Close the Peak inspiratory Pressure valve by turning the knob fully clockwise.

  2. Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20
    • Adjust the Max Pressure Relief knob clockwise or counterclockwise until manometer reads 40 cmH20

  3. Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.
    • Turn the Peak Inspiratory Pressure knob counterclockwise so the manometer reads 20 cmH20 and shut off the gas flow.

Conclusione

Alternate settings for the Max Pressure Relief should be made in accordance with hospital protocol.