Salta al contenuto principale

Hamilton-T1 Oxygen Cells Replacement

Attrezzature medicali
Né iFixit né questa guida di riparazione sono approvati dal produttore del dispositivo medicale o affiliati ad esso.

Cosa ti serve

  1. Hamilton-T1 Oxygen Cells Replacement, Oxygen Cells: passo 1, immagine 1 di 1
    • Flip device around until the back panel is exposed.

  2. Hamilton-T1 Oxygen Cells Replacement: passo 2, immagine 1 di 1
    • Remove the screw in the bottom center of the panel, and then take the panel off.

  3. Hamilton-T1 Oxygen Cells Replacement: passo 3, immagine 1 di 1
    • Disconnect the oxygen cell by removing the tube.

    • Then twist the cell counter-clockwise and pull to remove the oxygen cell.

    • Install the new oxygen cell and put the back cover on.

    • After the cell is installed. Run the oxygen cell calibration.

    • Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.

Conclusione

After the installation and calibration of the oxygen cell. The Hamilton-T1 Ventilator is ready to operate.