    • Remove the ribbon cable at the connection on the oxygen blender control board.

  2. Bear 1000 Air and O2 Blender Replacement: passo 2, immagine 1 di 1
    • Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.

  3. Bear 1000 Air and O2 Blender Replacement: passo 3, immagine 1 di 1
    • The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.

Conclusione

To reassemble your device, follow these instructions in reverse order. Perform an OVP test to check proper installment.

NOTE: do not interchange control boards with non-socketed ones!