Salta al contenuto principale

Autosampler Battery Replacement & Voltage Check (For Old & New Versions)

Cosa ti serve

  1. Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Disconnect the batteries completely: passo 1, immagine 1 di 3 Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Disconnect the batteries completely: passo 1, immagine 2 di 3 Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Disconnect the batteries completely: passo 1, immagine 3 di 3
    • 1. Disconnect the plug to the solar panel (yellow circle).

    • Then, grabbing the green connector and pull the battery connector from the solar charge controller (blue arrow).

    • This will be done inside the node. It was removed from the node to make it easier to see for the tutorial.

    • 2. Disconnect the plugs for the autosampler battery.

    • 3. Disconnect the connector that plugs into the battery port of the sampler.

    • If disconnect battery before solar panel, this can damage the solar charger!

  2. Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Checking battery voltages: passo 2, immagine 1 di 3 Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Checking battery voltages: passo 2, immagine 2 di 3 Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Checking battery voltages: passo 2, immagine 3 di 3
    • 4. Make sure multimeter is measuring in DC current (red boxes).

    • Make sure multimeter is measuring in Volts (blue box).

    • 5. Check the voltage of the lead acid autosampler battery. Fully charged lead acid battery should be around 13-14 V.

    • Remember: Red to red and black to black!

    • 6. Check the voltage of the lithium ion node battery. Fully charged lithium ion battery should around 4 V.

  3. Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Reconnect the batteries: passo 3, immagine 1 di 3 Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Reconnect the batteries: passo 3, immagine 2 di 3 Autosampler Battery Replacement & Voltage Check  (For Old & New Versions), Reconnect the batteries: passo 3, immagine 3 di 3
    • 7. First reconnect the autosampler battery (black box).

    • 8. Then reconnect lithium ion battery in the node (blue box).

    • Remember to plug in the other connector into the autosampler.

    • 9. Secure solar charge controller to the side of the enclosure using velcro (red box).

    • Then plug in the solar panel last (orange box).

Linea Traguardo

Altre 3 persone hanno completato questa guida.

Brooke Mason

Membro da: 05/29/19

2.928 Reputazione

17 Guide realizzate

Team

Open-Storm Membro di Open-Storm

Community

14 Membri

24 Guide realizzate

0 Commenti

Aggiungi Commento

Visualizza Statistiche:

Ultime 24 Ore: 0

Ultimi 7 Giorni: 0

Ultimi 30 Giorni: 0

Tutti i Tempi: 150